Article View: sci.techniques.microscopy
Article #32Re: UV Power supplies & electronics?
From: Gerald W. Gordon
Date: Mon, 13 Jul 1998 00:00
Date: Mon, 13 Jul 1998 00:00
47 lines
2977 bytes
2977 bytes
Our standard solution to this problem is to put the arc lamp power supplies on a separate circuit and use a surge/EMI suppressor on the line to the arc lamp power supplies. This prevents almost all the spike related problems so I presume that most of the noise is transmitted through the power line. However there is a stepper motor controller which sometimes gets confused when a lamp is ignited so it seem possible that the noise can be transmitted through the air. This is further supported by an incident many years ago when an early battery powered digital multimeter was fried when a Hg lamp was ignited nearby. I would expect that Xe lamp ignition makes a bigger noise pulse because Xe is harder to ionize (but a helper gas may be present in small quantities to get the arc started). Jerry gordon@med.unc.edu Paul Muhlrad wrote: > > (sorry if this appears twice; thought I sent it, but I don't see it on the newsgroups) > > I have some questions concerning the prudent use of UV (e.g. > mercury/xenon lamp) power supplies. We are setting up a new microscope > that is equipped with an Opti-Quip Model 1600 Power Supply powering a > 200 watt mercury/xenon lamp. I'm concerned about warnings I've read > about turning off all other electrical equipment before firing up the > lamp (because of a strong electromagnetic pulse induced when the lamp > ignites). We plan to follow this advice and turn off all the > electronics in the immediate vicinity of the power supply , but how far > (literally) do we have to go in shutting down electronics in the room. > For instance, must we turn off a computer monitor that is ~8 feet or a > computer that is ~12 feet away from the Opti-Quip--what is a safe > distance? Does it matter whether or not the electronics are on the same > power circuit as the power supply? Does the EMP originate from the > power supply or the lamp? What other precautions should we be taking? > (or am I being paranoid?). Thanks in advance for whatever advice anyone > can provide. > > Paul > -- > Paul Muhlrad > University of Arizona, Department of Molecular and Cellular Biology > 415 Life Sciences South, Tucson, AZ 85721 > > pmuhlrad@u.arizona.edu | (520) 621-9135 | fax (520) 621-3709 > http://www.mcb.arizona.edu/wardlab
Technical Details
Message-ID:
<35AA16CF.33D7@med.unc.edu>Path:
arc1-ca.pugleaf.net!unknown!not-for-mailReferences:
<35A687F8.AF04BC1D@u.arizona.edu>Show Full Headers
Complete Article Headers
X-Google-Language: ENGLISH,ASCII-7-bit X-Google-Thread: 10e1ff,831f5263691b6c94 X-Google-Attributes: gid10e1ff,public X-Google-Thread: 106237,831f5263691b6c94 X-Google-Attributes: gid106237,public From: "Gerald W. Gordon" <gordon@med.unc.edu> Subject: Re: UV Power supplies & electronics? Date: 1998/07/13 Message-ID: <35AA16CF.33D7@med.unc.edu> X-Deja-AN: 371028839 Content-Transfer-Encoding: 7bit References: <35A687F8.AF04BC1D@u.arizona.edu> Content-Type: text/plain; charset=us-ascii Organization: The University of North Carolina at Chapel Hill Mime-Version: 1.0 Newsgroups: sci.techniques.microscopy,bionet.molbio.proteins.fluorescent