Dominic-Luc Webb wrote:
> I have worked for some time with fluroescence polarization (FP)
> in studies of drug-protein binding. My drug, which is
> hydrophobic, typically has a fluorescent adduct and the
> solution has 0.01% Tween 20 to keep down non-specific binding
> to test tubes, multiwell plates, etc. Assay usually works well.
> 
> I usually do a check with a series of glycerol concentrations
> to ensure that FP is working OK. I expect at high glycerol
> concentration, the higher viscosity lowers molecular rotation
> which is detected as an increase in the amount of polarized
> light in the static (parallel) channel. This usually works
> fine, fluorescein for instance. I recently have found some
> labelled drugs that show no increase in polarization, and
> even a small decrease (about 20 mP) even when adding glycerol
> to 50%. Again, Tween 20 is 0.01% throughout. Fluorescence
> intensity remains relatively constant with only a slight
> parallel increase in both channels. The underlying physics of
> FP always seemed pretty straightforward, but now I am
> confronted with such data that has been reproduced several
> times and it is not clear how this came about.
> 
> Strangely, when a protein that is known to be a drug target is
> present, there is a strong increase in polarization (100-300 mP).

That is what I'd expect.  Binding to a large molecule (i.e., 
the protein) reduces the rotation rate of the labeled drug, 
thus increasing the polarization.

> The glycerol data has created an uncertainty about negative
> data for test proteins that I am screening for drug binding.
> 
> Is there anyone in here that might have some hints how this
> lack of change in polarization in glycerol might have come
> about, maybe some suggestions how to troubleshoot this?

It could be that the change of solvent to 50% glycerol 
reduces the fluorescence lifetime, thus increasing the 
measured polarization.

> 
> Dominic
> 
Technical Details
Message-ID: <049Mi.11183$bV2.6674@trndny02>
Path: arc1-ca.pugleaf.net!g2news2.google.com!news1.google.com!border1.nntp.dca.giganews.com!nntp.giganews.com!wns11feed!worldnet.att.net!199.45.49.37!cyclone1.gnilink.net!spamkiller2.gnilink.net!gnilink.net!trndny02.POSTED!9fcca4b5!not-for-mail
References: <Pine.LNX.4.44.0710011540080.19573-100000@uno.canit.se>

Show Full Headers
Complete Article Headers
X-Google-Thread: fe222,d7292fc16d0f3513
X-Google-Thread: fd9c4,d7292fc16d0f3513
X-Google-Attributes: gidfe222,gidfd9c4,public,usenet
X-Google-Language: ENGLISH,ASCII-7-bit
Path: arc1-ca.pugleaf.net!g2news2.google.com!news1.google.com!border1.nntp.dca.giganews.com!nntp.giganews.com!wns11feed!worldnet.att.net!199.45.49.37!cyclone1.gnilink.net!spamkiller2.gnilink.net!gnilink.net!trndny02.POSTED!9fcca4b5!not-for-mail
From: Marvin <physchem@verizon.net>
User-Agent: Thunderbird 2.0.0.6 (Windows/20070728)
MIME-Version: 1.0
Newsgroups: sci.techniques.spectroscopy,bionet.molbio.proteins
Subject: Re: Drug binding: When fluorescence polarization fails
References: <Pine.LNX.4.44.0710011540080.19573-100000@uno.canit.se>
In-Reply-To: <Pine.LNX.4.44.0710011540080.19573-100000@uno.canit.se>
Content-Type: text/plain; charset=ISO-8859-1; format=flowed
Content-Transfer-Encoding: 7bit
X-Antivirus: avast! (VPS 000777-4, 09/30/2007), Outbound message
X-Antivirus-Status: Clean
Lines: 43
Message-ID: <049Mi.11183$bV2.6674@trndny02>
Date: Mon, 01 Oct 2007 16:17:00 GMT
NNTP-Posting-Host: 71.167.164.128
X-Complaints-To: abuse@verizon.net
X-Trace: trndny02 1191255420 71.167.164.128 (Mon, 01 Oct 2007 12:17:00 EDT)
NNTP-Posting-Date: Mon, 01 Oct 2007 12:17:00 EDT