Re: Low temp icing problems

#271
From: Shekhar Mande
Date: Sun, 08 Oct 1995 00:00
12 lines
1460 bytes
One of the possibilities could be (which we accidentally encountered during
our runs)- is when the cryo-system was brought down by a user, all the liquid
N2 was not drained from the smaller dewar of the MSC system.  Due to this
the nozzle sucks in some water, which remains in the spiral tube.  If another
user starts the cryo-system next day or a day later, the water inside the
spiral freezes, causing rapid changes in the temperature of the cryo-flow, and 
some ice also falls from the nozzle onto the xtal.  The remedy is ofcourse
to dismantle the smaller dewar, and heat the spiral tube, which would evaporate
residual water in the tube and then start the gas flow before inserting
the spiral tube into the dewar again.

Shekhar
Technical Details
Message-ID: <459ab2$58e@nntp5.u.washington.edu>
Path: arc1-ca.pugleaf.net!unknown!not-for-mail
References: <450e7f$bal@prof.ese-metz.fr> <45149n$mjm@cnn.Princeton.EDU>

Show Full Headers
Complete Article Headers
X-Google-Language: ENGLISH,ASCII-7-bit
X-Google-Thread: fcf0b,a447db9b7c6c638d
X-Google-Attributes: gidfcf0b,public
X-Google-Thread: f804b,8e2922d4f4550687
X-Google-Attributes: gidf804b,public
From: Shekhar Mande <shekhar@gouda.bmsc.washington.edu>
Subject: Re: Low temp icing problems
Date: 1995/10/08
Message-ID: <459ab2$58e@nntp5.u.washington.edu>
X-Deja-AN: 117228365
references: <450e7f$bal@prof.ese-metz.fr> <45149n$mjm@cnn.Princeton.EDU>
content-type: text/plain; charset=us-ascii
organization: University of Washington
x-url: news:45149n$mjm@cnn.Princeton.EDU
mime-version: 1.0
newsgroups: sci.techniques.xtallography,bionet.xtallography
x-mailer: Mozilla 1.1N (X11; I; OSF1 V3.0 alpha)