Re: Double label florescence

#7736
From: Milton Charlton
Date: Sun, 28 May 1995 00:00
18 lines
1407 bytes
We found Bodipy was a little better than FITC for co-labelling with Texas Red.
There is less bleed into the red filter with Bodipy.  This using standard
Biorad 600 Ar-ion laser at 514nm.
When bleed can not be defeated, take advantage of the differntial bleaching
rate of Texas red and Bodipy/FITC.  We published a graph of differntial
bleaching with a real double labelled specimaec in
Neuron, 5, pg 773-779, 1990. " Strategic location of calcium channels at transmi
tter
release sites of frog neuromuscular junction".  The differential bleaching allow
ed us to
determine that bleed of the green signal through the red filter was not
a problem.  Using zoom we bleached a small box then zoomed out and took another
picture.  We could see the green bleached out in  the box but intact elsewhere.
The Texas Red stain o Ca channels was not affected in the area where green
had been bleached out.

Milton Charlton
University of Toronto
Technical Details
Message-ID: <9505281540.AA16033@spine.med.utoronto.ca>
Path: arc1-ca.pugleaf.net!unknown!not-for-mail

Show Full Headers
Complete Article Headers
X-Google-Language: ENGLISH,ASCII-7-bit
X-Google-Thread: 114d69,907ce7b26a228ae2
X-Google-Attributes: gid114d69,public
From: Milton Charlton <milton@SPINE.MED.UTORONTO.CA>
Subject: Re: Double label florescence
Date: 1995/05/28
Message-ID: <9505281540.AA16033@spine.med.utoronto.ca>
X-Deja-AN: 103430295
sender: Confocal Microscopy List <CONFOCAL@UBVM.CC.BUFFALO.EDU>
comments: Gated by NETNEWS@AUVM.AMERICAN.EDU
newsgroups: bit.listserv.confocal